A Dialysis Bloodline Can Detach Mid-Treatment: The Alert You Need to Review This Week

Certain B. Braun bloodline lots may detach during dialysis. We also review a HIT assay recall and updated instructions for Medline beds.

Hemodialysis bloodline connected to a hospital machine, with the connector highlighted in orange as a potential risk point during treatment.

Editorial

This week, we focus on two recalls that require healthcare organizations to identify specific products: certain B. Braun bloodline lots for the Dialog+ system and four lots of the HemosIL AcuStar assay used to detect antibodies associated with heparin-induced thrombocytopenia. We also revisit two issues previously covered by Bisturí Digital: Medline beds and StatStrip Gen 2.0 glucose meters.

B. Braun: Three Bloodline Lots Must Be Quarantined

The FDA issued an early alert concerning B. Braun’s Streamline Bloodline Set for Dialog+, model SL-2010M2096. A required solvent application step was not properly completed during assembly, potentially resulting in an inadequately bonded dialyzer connector.

The connector could detach or leak during priming or hemodialysis. This may cause blood loss, clotting or occlusion of the extracorporeal circuit, treatment interruption, or air entering the blood circuit. In extremely rare circumstances, air that passes the machine’s detection system could result in an air embolism.

The affected lots are:

  • A2600099
  • A2600128
  • A2600136

As of October 10, 2026, B. Braun had reported no serious injuries or deaths associated with this issue.

What to review: confirm the model and lot number of the bloodlines available in your hemodialysis unit. Affected products should be removed from use, quarantined, and returned to the manufacturer according to the recall instructions. Inform all relevant personnel and document the inventory reconciliation.

Read the FDA alert concerning B. Braun

HemosIL AcuStar: Negative Bias May Conceal a HIT Case

Instrumentation Laboratories, a Werfen company, recalled four lots of the HemosIL AcuStar HIT-IgG(PF4-H) assay. The FDA has identified this as the most serious type of recall.

The manufacturer confirmed performance variability involving a negative bias of 20% to 35% that may not be reliably detected through routine calibration and quality control. A sample could fall below the assay cutoff and be reported as negative even when the patient has antibodies associated with heparin-induced thrombocytopenia. Delayed recognition of HIT and continued exposure to heparin may increase the risk of thrombotic complications.

The affected lots are:

  • B40005 and B38062, part number 0009800127
  • B40006 and B38063, part number 0009802028

As of September 3, Werfen had reported no serious injuries or deaths associated with the issue.

What to review: identify and destroy affected lots according to your facility’s procedures, document their destruction on the manufacturer’s response form, and establish an alternative method for HIT antibody testing. Decisions about reviewing previously reported results, following up with patients, or repeating tests should be made by the laboratory, medical director, and treating clinical team.

Read the FDA recall notice for HemosIL AcuStar

Follow-Up: Medline Allows Temporary Use Under Updated Instructions

The Medline bed alert is not new to our readers, but the current instructions require careful interpretation. Affected beds, rails, mattresses, and accessories may remain in use while they are repaired or replaced, provided the manufacturer’s updated measures are followed.

Medline instructs users to keep beds unplugged from electrical outlets. A bed should be plugged in only when the patient needs to be repositioned and unplugged when repositioning is complete. Extension cords must not be used and, whenever possible, a second person should assist with operating or moving the bed. Do not attempt to repair or modify a bed outside a Medline-authorized process.

The alert covers the risks of overheating, smoke, sparks, fire, electrical shock, and entrapment. As of October 20, Medline had reported two deaths and 16 serious injuries associated with these issues.

Read the updated FDA instructions for Medline beds

Follow-Up: StatStrip Gen 2.0 Is Still Awaiting an Update

On October 30, the FDA clarified the recommended actions for StatStrip Gen 2.0 hospital glucose meters, models 63685 and 66739. We covered this subject in previous editions and are not presenting it as a new alert.

If an alternative meter is available, the recommendation is to stop using the affected device until the software update has been installed. When no alternative is available, the mitigations should remain in place: validate patient identification in NovaNet when ADT integration is used, automatically log the operator out after each test, and verify the patient’s identity on the screen both after scanning and when the result appears. The defect affects the assignment of the result, not the analytical glucose measurement.

Read the FDA update for StatStrip Gen 2.0

In Brief: Two Advisories for Hospital Infrastructure

  • Red Hat, CVE-2026-98432: a critical vulnerability in the Linux kernel’s IPv4 subsystem may allow an unauthenticated remote attacker to cause memory corruption, a system crash, or remote code execution. Red Hat Enterprise Linux 8, 9, and 10 are affected; some OpenShift products are also affected or still under investigation. Inventory the installed versions and apply the correction specified by Red Hat for each product.
  • Cisco ASA and FTD, CVE-2026-76418: a high-severity vulnerability in the Remote Access SSL VPN service may allow an unauthenticated remote attacker to reload the device and cause a denial-of-service condition. Cisco provides no workaround; review the affected releases and upgrade to a fixed version. The official advisory does not state that the vulnerability is being actively exploited.

Closing Thoughts

This week’s priority is not to review “all equipment” in a general sense, but to identify exact models, part numbers, and lots. In dialysis, separate the affected B. Braun bloodlines before the next treatment. In the laboratory, remove the HemosIL lots and agree on an alternative method with the clinical team. For the previously known issues, confirm that Medline and StatStrip are being managed according to the latest instructions.

Share this edition with clinical engineering, nephrology, the clinical laboratory, nursing, and information technology. A short, well-documented review may prevent a treatment interruption or a clinical decision based on incorrect information.

Until next week.

Roberto Manzo

The analysis and prioritization expressed in Bisturí Digital are the editorial responsibility of Roberto Manzo. This newsletter does not replace manufacturer instructions, clinical assessment, or regulatory advice specific to your organization.